+ | === Coulbourn (feeder, valve, lever etc.) wiring === | ||
+ | {{:guides:coulbourn_wiring.docx|Coulbourn wiring doc}} | ||
=== RR2 === | === RR2 === | ||
[[guides:homemade:RR2commutator|RR2Commutator]] | [[guides:homemade:RR2commutator|RR2Commutator]] | ||
+ | |||
+ | === Neuronexus Buzsaki 64L === | ||
+ | |||
+ | ==Probe Site Maps== | ||
+ | [[http://www.neuronexustech.com/images/Electrode%20Site%20Map/A64%20Maps.pdf]] | ||
+ | |||
+ | ==Neuronexus Impedance Charts== | ||
+ | [[https://www.dropbox.com/s/prq5ze6qm756q04/2013_03_25_19_36_28.pdf]] | ||
+ | |||
+ | ==Si Probe Plating== | ||
+ | |||
+ | Use SIFCO Non-cyanide gold solution (Part Number 5355). | ||
+ | |||
+ | For plating parameters, use the GUI for the Nano-Z. Current is set | ||
+ | for 1.5 uA. Target impedance is 400 Kohm at 1004 Hz for a max of 5 | ||
+ | runs. Use 1 second intervals with 1 second pauses. Use 'Match | ||
+ | Impedances' mode. | ||
+ | |||
+ | |||
+ |