Line 8: Line 8:
 === Neuronexus Buzsaki 64L === === Neuronexus Buzsaki 64L ===
  
-Probe Site Maps   +==Probe Site Maps==   
 [[http://​www.neuronexustech.com/​images/​Electrode%20Site%20Map/​A64%20Maps.pdf]] [[http://​www.neuronexustech.com/​images/​Electrode%20Site%20Map/​A64%20Maps.pdf]]
  
-Neuronexus Impedance Charts +==Neuronexus Impedance Charts==   
 [[https://​www.dropbox.com/​s/​prq5ze6qm756q04/​2013_03_25_19_36_28.pdf]] [[https://​www.dropbox.com/​s/​prq5ze6qm756q04/​2013_03_25_19_36_28.pdf]]
  
-Si Probe Plating+==Si Probe Plating==
  
-We use SIFCO Non-cyanide gold solution (Part Number 5355).+Use SIFCO Non-cyanide gold solution (Part Number 5355).
  
-For plating parameters, ​we use the GUI for the Nano-Z. ​ Current is set+For plating parameters, use the GUI for the Nano-Z. ​ Current is set
 for 1.5 uA.  Target impedance is 400 Kohm at 1004 Hz for a max of 5 for 1.5 uA.  Target impedance is 400 Kohm at 1004 Hz for a max of 5
-runs.  ​We use 1 second intervals with 1 second pauses.  ​We use '​Match +runs.  ​Use 1 second intervals with 1 second pauses.  ​Use '​Match 
-Impedances'​ mode.  Sorry I don't have a script for this.+Impedances'​ mode. 
 + 
 + 
 + 